China Instrument for testing calcium oxide in stone powder and gravel - China Supplier China Instrument for testing calcium oxide in stone powder and gravel - China Supplier China Instrument for testing calcium oxide in stone powder and gravel - China Supplier China Instrument for testing calcium oxide in stone powder and gravel - China Supplier China Instrument for testing calcium oxide in stone powder and gravel - China Supplier

Instrument for testing calcium oxide in stone powder and gravel

Price:Negotiable
Industry Category: Chemicals/Inorganic Salts/Other Inorganic Salts
Product Category:
Brand: 標(biāo)創(chuàng)儀器
Spec: DT-8000A


Contact Info
  • Add:鶴壁市山城區(qū)朝陽街東段, Zip: 458000
  • Contact: 周經(jīng)理
  • Tel:13503920302
  • Email:

Other Products

Description
Additional Information

Instrument for Testing Calcium Oxide in Stone and Stone Powder Instrument for Testing Calcium Oxide in Stone and Stone Powder - Hebi Biaochuang Instrument Manufacturing Co., Ltd.


First, prepare the calibration samples with known accurate results, number them, and prepare them for pressing into pellets.

1. After turning on the instrument, pull out the slide plate and press the Enter key. The preheating time will start counting down. Within 30 minutes of startup, enter the "Main Menu" program. On the display screen [3-2], the message "Preheating仍需 * * minutes" will appear at the bottom. After 30 minutes of preheating, this message will disappear, indicating that the instrument has completed preheating. If, 30 minutes after startup, you enter the "Main Menu" [3-2] screen and the message "Preheating 30 minutes" does not appear at the bottom, it also means the instrument has been preheated.

1. Startup interface: Click (3) System Settings and enter the password to proceed;

1. Start by pressing (3) System Settings → Enter the password six 1s (111111) to enter the calibration program;

2. Press (1) Clear Data to ensure previous data does not affect this calibration;

3. Press (2) Measure Calibration Sample. Pull the slide plate out completely, place the *** numbered calibration sample, then push the slide plate back in completely. Press (1) Start. After the countdown ends, the buzzer will emit a continuous "beep beep beep..." sound, and the display will show "(Please pull out the slide plate)". At this point, pull the slide plate out completely. The prompt message on the screen will disappear, the sound will automatically stop, and the instrument will begin measuring the reference sample. The countdown will be displayed in the middle of the screen. After it ends, press (0) to return. Repeat this process for the second sample: place it, push the slide plate in, press 1 to start, pull out the slide plate after the countdown ends, wait for completion, then place the third sample.


After all calibration samples have been calibrated, press 0 to return, then press 3 and 1 to view the CaO count data. Send the images to the company's technical personnel for guidance on the next steps. If the calibration sample counts increase with higher content, the next step is to input the chemical content.

4. Input Chemical Content: Enter the content of the calibration data from low to high. This is for calculating the K and b values in the next step;

5. Calculate K and b Values: After inputting the chemical content, proceed to calculate the K and b values. Press 0 to return to the calibration interface →

Press 5 to calculate K and b values → Press to calculate K and b values → 1. Enter the working curve number 1 and press Enter.

Prompt: Press 2 to select the K and b value calculation item → Then press 1 for CaO → Press 0 to return → 3 K and b value calculation;

II. Content Measurement

1. After the instrument is calibrated, it can be used to measure the content of the samples to be tested.

2. Sampling Situation. At the very bottom of the display screen, the prompt messages "Sampling①②③④⑤" will also disappear. When the word "Sampling" appears, the sample slide plate must be fully pulled out so that the slide plate is in the reference sample measurement position. The lower left corner of the display will show "Reference Sample". Sampling can be completed in approximately 1.5 minutes.

3. Select the Correct Working Curve Number. Use the working curve number where the K and b values were input during calibration to accurately analyze the samples to be tested corresponding to the calibration samples. The working curve is displayed in the upper left corner of the screen, with a total of 1-9 curves. They cycle through by pressing the [ ] key repeatedly. Finally, select the working curve number where the K and b values were stored during calibration.

Note: After shutting down or a power outage, when the instrument is turned on again, it will be in the initial state with the working curve number set to 1. Pay special attention to whether the working curve number is selected correctly at this time.

Industry Category Chemicals/Inorganic Salts/Other Inorganic Salts
Product Category
Brand: 標(biāo)創(chuàng)儀器