尤物视频污,99热99热99超碰,精品久久久久女少妇,少妇精品无码,日B视屏久久香色,亚洲美女午夜福利av,国产美女被操,国产九九精品,国产久久色情

China GZX92A Insulation Resistance Meter Calibration Device - China Supplier
China GZX92A Insulation Resistance Meter Calibration Device - China Supplier

GZX92A Insulation Resistance Meter Calibration Device

Price:Negotiable
Industry Category:
Product Category:
Brand: 深圳世紀經典
Spec: GZX92A型絕緣電阻表檢定裝置


Contact Info
  • Add:深圳市龍崗區(qū)銷售熱線:15914142916, Zip: 518116
  • Contact: 馮先生
  • Tel:15914142916
  • Email:186jl@163.com

Other Products

Description
Additional Information

I. Overview

The insulation resistance meter (megohmmeter) is a nationally mandatory verification measuring instrument, with its verification regulation being JJG622-89. The main standard devices of the verification device are the high-voltage high-resistance box and the high-voltage voltmeter. The GZX92A insulation resistance meter verification device integrates the high-voltage high-resistance box (high-resistance output part) and the high-voltage digital voltmeter (voltage input part) into one unit.

The measurement upper limit of the resistance output part in the verification device is 200 GΩ, with a nominal working voltage of 5 kV. The device is designed with independent leakage terminals and grounding terminals, showing no significant instability, short-circuit, or open-circuit phenomena during measurement. It meets the requirements of the JJG622-97 insulation resistance meter verification regulation and the JJG166-93 DC resistor verification regulation. It can be used for the verification of various models of pointer-type insulation resistance meters.

The resistance output part of the verification device uses a patented rolling high-voltage switch, which can still meet the verification regulation requirements after 100,000 uses. It has a long technical and mechanical lifespan. The nominal measurement voltage of the voltage measurement part in the verification device is 5 kV, with an input impedance of ≥10 GΩ. This device can measure the open-circuit voltage, median voltage, and peak voltage of the megohmmeter.

II. Main Technical Parameters

1. Resistance Output (High-Voltage High-Resistance Box) Part

1.1 Accuracy Class and Working Voltage (Current) of the Resistance Output Part

Resistance Value

100 GΩ

×10 GΩ

×1 GΩ

×100 MΩ

×10 MΩ

×1 MΩ

Accuracy Class

5

5

2

1

0.5

0.2

Nominal Voltage

5000 V

5000 V

5000 V

5000 V

2500 V

1000 V

Resistance Value

×100 kΩ

×10 kΩ

×1 kΩ

×100 Ω

Accuracy Class

0.2

0.2

0.2

0.2

Nominal Current

1 mA

8 mA

20 mA

50 mA

1.2 Adjustment Range: 100 Ω–200 GΩ, with a fine adjustment of 100 Ω

1.3 Operating Environmental Conditions

1.3.1 Reference Temperature Range: 20–25°C

1.3.2 Nominal Operating Temperature Range: 18–28°C

1.3.3 Reference Humidity Range: 40–60%

1.3.4 Nominal Operating Humidity Range: 25–75%

1.4 Resistance Variation Limit: Under reference conditions, the variation caused by a change in a single influencing factor

Influencing Factor

Nominal Operating Range

Permissible Variation

Ambient Temperature

18–28°C

a/2 %

Relative Humidity

25%–75%

≤10 GΩ a/2 %; >10 GΩ a%

Operating Voltage

(1–1/5) Nominal Voltage

≤10 GΩ a/2 %; >10 GΩ a%

★ a is the accuracy class of the resistance dial in the resistance output part of the verification device (a values differ for each resistance dial).

1.5 Insulation Resistance: The insulation resistance between the circuit of the resistance output part of the verification device and any other external metal without electrical connection. The resistance value measured at the nominal voltage is not less than 5 TΩ.

1.6 Insulation Strength: Between the circuit of the resistance output part of the verification device and the reference grounding point for testing, it should withstand an actual sinusoidal AC voltage of 11 kV at a frequency of 45–65 Hz for 1 minute without breakdown or flashover.

1.7 The residual resistance at the output terminals of the resistance output part of the verification device should be <0.1 Ω, with a variation of <0.01 Ω.

1.8 Dimensions: 442 mm × 270 mm × 145 mm

1.9 Weight: <5 kg

2. Voltage Measurement (High-Voltage DC Digital Voltmeter) Part

2.1 Measurement Range: 0–5500 V

2.2 Accuracy: ±(1% of reading ±1 digit)

2.3 Power Supply: DC 9 V

2.4 Input Resistance: ≥10 GΩ

2.5 Display: 4.5-digit, LCD display

2.6 The peak voltage measurement circuit meets the requirements of Figure 3 in the JJG622-97 insulation resistance meter verification regulation.


Industry Category
Product Category
Brand: 深圳世紀經典
Spec: GZX92A型絕緣電阻表檢定裝置
Stock:
Origin: China / Guangdong / Shenshi
About Toocle.com - Partner Programme - Old Version
Copyright ? Toocle.com. All Rights Reserved.
(浙)-經營性-2023-0192
ChatGlobal Chat Now 屏南县| 藁城市| 扎囊县| 响水县| 搜索| 资中县| 宜宾市| 濉溪县| 东乡族自治县| 宁陵县| 板桥市| 五华县| 鄂托克前旗| 永修县| 隆尧县| 九台市| 泌阳县| 饶平县| 冕宁县| 乌海市| 双江| 丹凤县| 白玉县| 应用必备| 闽侯县| 巴林右旗| 柘城县| 大庆市| 仙居县| 长岛县| 台南县| 南充市| 沈阳市| 天水市| 江山市| 南华县| 齐齐哈尔市| 阿克苏市| 奎屯市| 张家口市| 武胜县|